    DM500 ultrasound density sensor mixer milk liquid ultrasonic density meter high stability liquid compatible with 304L

    1 Overview The DM500 is one type of liquid density sensors which calculate the density from ultrasonic wave speed in the liquid. As it is based on the principle of speed related to the densi

    1. Detailed information

    1 Overview

     

    The DM500 is one type of liquid density sensors which calculate the density from ultrasonic wave speed in the liquid.

     

    As it is based on the principle of speed related to the density of media. So it is applicable for the liquid which is with known value of referenced density. It is compact design with small dimension and it is easy to install. It is not necessary to be vertical in the liquid. Only if it is in the liquid, it is enough.

     

    Furthermore, it is applicable for customized design, such as different mechanic structure, measurement range, and other OEM/ODM business.

     

     

     

     

    2 Specification

     

    Measurement

    0.1~2g/cm3



    Range







    Resolution

    0.01g/cm3




    Drift per year

    ±0.5%FS




    Power Supply

    12-24V DC



    Output Signal

    RS232/RS485



    Working

    -40 ~ 85, 595%RH


    Environment






    Protection Rank

    IP68  for

    the

    transducer,  IP66  for


    controller box






    Wireless

    Optional

    for

    zigbee,  wifi,  GPRS,


    3G(UMTS).






     

     

    3 Principle

     

    For the DM500, the sensor directly measure the ultrasonic wave speed v1 in current media.Then calculate the rate of v1 to reference media ultrasonic wave speed v0.Finally draw out the density of current media p1 based on the referenced media density p0. Formula is

     

    p1:p0=(v1:v0)^2.

    4 Interface Definition

     

     

    Pin 1: Red Line: V+,

    Pin 2: Blue Line: RS485 A+/ RS232 TX

     

    Pin 3: Black Line: GND

    Pin 4: Yellow Line: RS485 B-/ RS232 RX

     

    If there is label of wire definition on the device, please act as the label said!

     

    5 Part List

     

    The package include two parts, one is the ultrasonic transducer which is need to be put into the liquid to be measured, the other is the controller box.



    6 Quick Start

     

    6.1 Install Density Reading Software

     

     

    https://yadi.sk/d/kQHpvuhz3JKDrg is the software for density configuration and reading. User can download to install it.

     

    Of course, user can send command according to the protocol to configure and read sensor. Not necessary to use the density configuration and reading software.

     

    6.2 Set Current Speed as reference speed

     

     

    Connect sensor with PC and power supply according to Part 4.

    Put ultrasonic transducer into liquid. And power on the sensor.

    Click button of set current speed as reference. The sensor will use the current speed as referenced speed v0. (See part 3 for the principle and meaning of v0 v1 p0 p1).

     

    If command send successfully, the sensor will reply the liquid density: xx.

    6.3 Set referenced density

     

     

    Input the density value for the referenced liquid. Click send button. The sensor will use the current media as referenced liquid density p0.

     

     

     

    6.4 Read density

     

     

    Put the ultrasonic transducer into the liquid to be measured, click read density button. The sensor will output the density of current liquid.

    Notes:

    Please wait for >=5 seconds after moving from referenced media to the media to be measured.

     

    If the ingredient of the liquid to measured is significantly different from the referenced liquid. The output of density maybe different from real data. E.g, 5% ethanol with water can be used as reference for media 20% or 50% ethanol. But 5% ethanol with water can not be used as reference to measure diesel with water.

     

    7 Protocol

     

    7.1 Com Port Setting

     

     

    Baudrate115200

     

    Data bit8


    Stop bit1

     

    Check SumNone

     

    • is reserved for this protocol, it is specially used to splitter for the parameter. In user data, it should not be used as parameter.

       

      7.2 Packet Format

       

       

       

       

    1

    2

    2

    n

    2

    2

    Header

    Command

    ID

    Content

    Checksum

    Tail

     

    Header/Tail

     

    Header 1 byte$ in ascii and in hex 0x24 for the request to sensor. And * in ascii and in hex 0x2A for the reply from sensor. Tail 2 byte, in hex 0x0D 0A.

     

    Command

     

    Refer to command list at Part 3. For command in reply is same as the command in request.

     

    ID

    01~FF

    Content

    0~32 bytes.

    Checksum

     

    Sum of bytes from header to id, then select the low 8 bit, finally convert to ASCII and show in Hex. E.g. If sum is 0x0356, then low 8 bit is 0x56, to ascii, the value is 56(0x35 0x36 in Hex).

    Example

     

    $RD011B Read Density

     

    In hex 0x 24 52 44 30 31 31 42 0D 0A

    0x 24 + 55 + 41 +30 + 31 = 0x 11B , the low byte is 0x1B, then checksum is 1B in ascii, in hex 0x31 42.

     

    Notes:

    For the automatic update version, the sensor automatically upload.

     

    7.3 Command List

     


    Command

    Function

    Scope





    Standard Version





    SN

    Set Calibration

    RS

    SS

    Set reference Speed

    RS




    SD

    Set reference Density

    RS




    SI

    I

    RS


    Set  D


    SL

    Set Liquid Type

    RS




    RD

    Read Density

    RS

     

    Notes: SMS: applicable for SMS message. RS: applicable for RS232/485. Wireless:applicable for Wifi/zigbee/GPRS/3G.

     

    7.4 Detailed Command

     

     

    For the DM500, the sensor directly measure the ultrasonic wave speed v1 in current media.Then calculate the rate of v1 to reference media ultrasonic wave speed v0.Finally draw out the density of current media p1 based on the referenced media density p0. Formula is

     

    p1:p0=(v1:v0)^2.

     

     

    7.4.1 Set Calibration

     

     

    Set current media as referenced media, so the speed in current media will be

     

    used as v0. The density for this current media will be referenced density. If

    not specify the value, it will be treated as 1. In another word, this command

    should be done at the reference media.

    Request from computer to sensor:

    ASCII :  $SN0126

    Hex: 24 53 4E 30 31 32 36 0D 0A

     

    Reply from sensor

     

    Liquid Speed: 1.519

     

     

     

    7.4.2 Set Reference Speed

     

     

    Set speed of reference media v0. Usually user do not know the reference

    speed, please us the Set Calibration command at the reference media and Set

    Reference Density command.

    Request from computer to sensor:

    ASCII : $SS0101.48056

     

    Hex : 24 53 53 30 31 30 31 2E 34 38 30 35 36 0D 0A

    Reply from sensor

    Liquid Speed: 1.480

     

    Unit is km/s.

     

    7.4.3 Set Reference Density

     

     

    Set reference density p0 value. Unit is g/cm^3. This command can be done in any media. In another version, it is not necessary to use this command in the referenced media.

     

    Request from computer to sensor:

    ASCII : $SD0101.1113E

    Hex : 24 53 44 30 31 30 31 2E 31 31 31 33 45 0D 0A

    Reply from sensor:

    Liquid Density:  1.111

    Unit is g/cm^3.

     

    7.4.4 Set ID

     

     

    Request from computer to sensor:

    ASCII : $SI0100000142

    Hex : 24 53 49 30 31 30 30 30 30 30 31 34 32 0D 0A

    Reply from sensor:

     

    *SI01OKOKOKEF

    01 is sensor id, 00 is filler to reach 6 bytes, 01 is the new id, id range is 00~FF.

     

    7.4.5 Set Liquid Type

     

     

    Request from computer to sensor:

    ASCII : $SL0100000347

    Hex : 24 53 4C 30 31 30 30 30 30 30 33 34 37 0D 0A

    Reply from sensor:

    *SL010000034D

    01 is sensor id, 00000 is filler to reach 6 bytes, 3 is the liquid type.

     

    1: water, 2: gasoline, 3: diesel

    7.4.6 Read Density

     

     

    Read current media density.

    Request from computer to sensor:

    ASCII : $RD011B

     

    Hex : 24 52 44 30 31 31 42 0D 0A

    Reply from sensor:

    ASCII: *RD0101.11143

    Hex: 2A 52 44 30 31 30 31 2E 31 31 31 34 33 0D 0A Density is 1.111, Unit is g/cm^3.

     

    For some version, it is automatic upload *RDxx.

     

     

    8 Naming Rules & Clarification

     

     

    DM500 Ultrasonic Density Sensor

     

    Output

     

    RS485

     

    CRS232 Wireless

     

    W Wifi

     

    Z zigbee

     

    G GPRS

     

    U 3G(UMTS)

     

    9 Video

     

    DM500 ultrasonic density sensor test with liquids https://www.youtube.com/watch?v=uz1Xwas3ZGs&t=96s

     

    DM500 Test in Beer

    https://www.youtube.com/watch?v=OWOpfj-5tPI&t=35s